Questions and Answers

Webinar 13

Q: Basic question - when I load .wiff file from MRM EXPT. I am not seeing the chromatogram - what am I doing wrong?

Ans: It may be that you do not have your targets configured correctly. They need to match up with the transitions measured in your .wiff file. If you didn't generate the .wiff file from a method exported from Skyline that can be a little more challenging. Consult the Existing and Quantitative Experiments tutorial. The best way to proceed would be to post a message to the support board with at least a screenshot. If we still need more information, we'll ask you to post your files so that we can have a closer look at what is going on.

Q: Wouldn't losses during protein digestion lead to LOWER amounts than expected, not higher?

Ans: It is possible. It would depend on the experimental design. One possible explanation Chris offered depended on when the heavy reference peptide standard was added to the sample. For instance, if it was added prior to digestion, that could cause non-specific cleavage degrading some of the standard peptides, which would cause ratios of light to heavy to be overestimated. 

Q: Does the multipoint calibration work with several replicates from the same standard (eg 3 points for 10fmol/ul, 3 points for 40fmol/ul...)

Ans: Yes. Skyline supports multiple replicates per standard concentration. That is actually shown near the end of the tutorial when the CPTAC Study 9 data set is downloaded and merged into a document containing 72 runs with a 10-point dilution series and 4 replicates of each concentration.

Q: Suppose I do not want all the daughter ions for the calibration curve.  Let’s say I need only 2.  How do I remove the others?

Ans: You could use Edit > Refine > Advanced to remove all but the 2 most intense. In version 3.7.1.xxxx and later, you can also leave the extra transitions in place as "qualitative" and turn off a new Quantitative property exposed on all transitions. You need to set this up using the Document Grid View designer, but once you have turned off the Quantitative property the affected transitions will no longer be included in any Total Area values and neither Group Comparisons nor Quantification.

Q: Would it be possible to generate calibration curve when heavy peptide concentration changes and light peptide concentration is the same?

Ans: Yes, that is possible in the Peptide Settings - Modifications tab by specifying that the reference standard should be considered the "Light" isotope label type.

Q: Would it be possible to use LOG-LOG linear regression fit?

Ans: A linear regression on LOG-LOG is currently not available, as it is not actually fitting a line to the data unless the slope is one. Skyline does offer a 1/(x * x) regression weighting, which some people use for a similar purpose to give more weight to lower concentrations and avoid letting the the highest concentrations dominate the regression. This seems to work best when the concentrations are mostly linear. As more standard concentration points fall below the LOD, they will throw off the calibration more because of the high weighting at the lower concentrations. The same would probably be true of LOG-LOG.