Title | | » | Need some way to show user FAIMS information in raw scan view, to demystify jagged chromatograms before IM filters are applied |
Assigned To | | » | Brian Pratt |
Notify | | » | Brendan MacLean (test) |
Type | | » | Defect |
Area | | » | Skyline |
Priority | | » | 3 |
Ion mobility library "Use Results" generally takes care of the problem of using FAIMS data in Skyline (where you get jagged chromatograms without IM filtering), but there's no way to inspect the values directly. (For other IM types we have the heatmap view, but FAIMS isn't really 3D - the CV changes at discrete RTs.)
We should show that CV value somewhere in the raw scan viewer. Exactly where is a design question...
1) On the same line as existing scan details? (not great, already pretty crowded)
2) Split scan details into two lines, and maybe move time down there as well? (my choice but Brendan doesn't love it)
3) Put the CV info on the upper left of graph area? (Brendan's choice, but I worry that it will interfere with peak annotation display)