Issue 489: Importing negative ion transitions from Masshunter (small molecules)

issues
Status:closed
Assigned To:Guest
Type:Defect
Area:Skyline
Priority:2
Milestone:3.6
Opened:2016-11-14 14:36 by melechco
Changed:2017-10-18 08:36 by Brendan MacLean
Resolved:2016-11-15 12:59 by Brian Pratt
Resolution:Fixed
Closed:2017-10-18 08:36 by Brendan MacLean
2016-11-14 14:36 melechco
Title»Importing negative ion transitions from Masshunter (small molecules)
Assigned To»Brendan MacLean
Type»Defect
Area»Skyline
Priority»2
Milestone»3.6
Skyline is not recognizing data acquired in negative mode (negative chemical ionization GC-MS/MS) from MassHunter. It is going to import only by setting positive charge to all transitions (all were acquired in negative mode).

At 8:31 PM:
Failed importing results file '\\fleuryfs\IPD\CROM\CG\20161114-02.D'.
This document contains only negative ion mode transitions, and the imported file contains only positive ion mode data so nothing can be loaded.


 
 20161114-01.D.zip

2016-11-14 14:47 Brian Pratt
Assigned ToBrendan MacLean»Brian Pratt

2016-11-15 12:59 Brian Pratt
resolve as Fixed
Statusopen»resolved
Fixed in 3_6 branch, merged to trunk. We were not properly noting the transition polarity for Agilent chromatogram data.

2017-10-18 08:36 Brendan MacLean
close
Statusresolved»closed
Assigned ToBrian Pratt»Guest