Can you send me the .raw files for those replicates?
Files which are less than 50MB can be attached to these support requests. You can always upload larger files here:
https://skyline.ms/files.url
It is indeed mysterious that the molecules in 242064 have MS1 chromatograms that cover different time ranges.
The usual cause of that is that the "Retention time filtering" setting on the Full Scan tab at "Settings > Transition Settings".
You have that set to "Use only scans within 5 minutes of MS/MS IDs". That setting could certainly cause you to have truncated chromatograms, but only if you had a spectral library with retention times in it, which you do not seem to have.
You could certainly try changing that to "Include all matching scans" and then telling Skyline to extract chromatograms again by going to "Edit > Manage Results" and pushing the "Reimport" button.
The other thing that could cause your MS1 chromatograms to be different lengths is that Skyline treats MS1 spectra with a m/z range less than 500 special to support SIM or boxcar data where MS1 spectra might be intended to be used for only the precursors that fall in the scan range. However, your dataset does not seem to have any spectra like that.
After I see you .raw files I will be able to watch Skyline extract chromatograms and I will probably be able to figure out what is going wrong.
-- Nick