Issue 370: Small Molecule CE Output not working

issues
Status:closed
Assigned To:Guest
Type:Defect
Area:Skyline
Priority:3
Milestone:4.1
Opened:2014-12-02 16:27 by Brian Pratt
Changed:2017-07-03 11:50 by Brian Pratt
Resolved:2017-07-03 11:49 by Brian Pratt
Resolution:Fixed
Closed:2017-07-03 11:50 by Brian Pratt
2014-12-02 16:27 Brian Pratt
Title»Small Molecule CE Output not working
Assigned To»bspratt
Type»Defect
Area»Skyline
Priority»3
Milestone»3.1
Will Thompson created a new request.    2014-11-20 07:01
HI Brian

I have been toying around with the new small molecule feature in Skyline daily and have noticed a bug for method export. I have built a new 'regression' for Collision energy (CE) for some small molecules, however when i export the .exp instrument method for our Waters Xevo TQS all of the collision energies in the method are set to zero. I have tried even with the factory 'default' CE profile (Waters Xevo) and the CE are still exported as zero. This obviously means the collision energies for each channel must be entered by hand, which is a bad bug to have. :) Thanks for taking a look.

I have attached the small molecule skyline document i am working with.

Cheers

Will
 
 3805_SmallMoleculeDemo_111714.sky.zip

2017-06-08 15:30 Brendan MacLean
Milestone3.1»4.1
Still a problem?

2017-07-03 11:49 Brian Pratt
resolve as Fixed
Statusopen»resolved
Corrected in SVN 8768.

2017-07-03 11:50 Brian Pratt
close
Statusresolved»closed
Assigned ToBrian Pratt»Guest