From David Creasy:
We mentioned two features in Distiller that could cause problems for Skyline:
1. The option to save the Fragment ions in MS/MS peak lists as MH+ values will presumably cause problems for Skyline:
[See image 1 attached]
2. The processing option to use the precursor charge as the maximum charge allowed for fragment ions will cause multiple peak lists to be written out for each ms-ms spectrum:
[See image 2 attached]
This can also be caused by clearing the option for a "Single peak list for multiple precursor charges" above. This shouldn't cause a problem for Skyline, but you will see the same (or nearly the same) spectrum appear multiple times. You can detect this because the same TITLE= line will appear multiple time. The difference in the spectra might be for example where a 3+ fragment is excluded from a spectrum where the pre-cursor is 2+
The value:
_MSMS.UsePrecursorAsMaxCharge = 0
appears near the top of the MGF file.